Why use a B-10 lined proportional counter vs. a fission chamber for neutron detection?
Each detector type offers unique characteristics which suit it to a particular application. B-10 lined proportional counters offer excellent sensitivity to thermal neutrons, allowing them to achieve sufficient count rates in low neutron fluxes. For example, this feature makes the use of B-10 proportional counters attractive as monitors during initial reactor criticality, where the increased detector sensitivity ensures that smaller, less expensive, start-up sources can be installed in the reactor core. Fission chambers, in contrast, have lower sensitivity to thermal neutrons but have superior resistance to gamma ray interference and offer the capability to operate over many decades of neutron flux. This feature makes fission chambers the ideal choice as wide range monitors of reactor power. The Photonis technical team is happy to work with customers to determine which detector type makes the most sense in each customer application.