With Phasics's unique QWLSI patented technology, test your spectral filters at their designed working wavelength as well as polarizing optics with an interferometric precision. Combining achromaticity and polarization independent measurements, Phasics standalone wavefront sensors and integrated test bench provide transmitted wavefront error (TWE) and reflected wavefront error (RWE) measurements of bandpass spectral filters as well as polarizing optics. Simply tune the probe wavelength to match the design wavelength of your optics and it's done!
Besoin de précisions ? Contactez nos experts !
Contactez-nous
Transmission Wavefront Error (TWE) measurement setups
Either using a Kaleo MultiWAVE fully integrated machine or a modular solution from Phasics, TWE measurements are done in a similar double-pass configuration as a standard Fizeau interferometer. Single-pass measurements can also be performed if required. First, a reference measurement is taken using a reference flat. Then, the filter is placed in the beam path to acquire the transmitted wavefront error. The wavelength can be adjusted either using a tunable source or multiple individual sources.
Reflected Wavefront Error (RWE) measurement setups
Similarly, RWE measurement can be performed using the Kaleo MultiWAVE machine or cutom-made optical test setup. First, a reference measurement is taken using reference optic, then the reference optic is replaced by the unit under test, and the reflected wavefront error is measured.